CVE-2025-41690
Endress+Hauser: Proline 10 Maintenance credentials may be exposed under certain conditions
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Fix
- Fixed in 01.00.06, 01.00.02
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
CISA's decision points: whether it is exploited, whether attacks can be automated, and how much control an attacker gets. Assessed (SSVC 2.0.3).What do these mean? →
- Exploitation: none
- Automatable: no
- Technical impact: total
Summary
A low-privileged attacker in bluetooth range may be able to access the password of a higher-privilege user (Maintenance) by viewing the device’s event log. This vulnerability could allow the Operator to authenticate as the Maintenance user, thereby gaining unauthorized access to sensitive configuration settings and the ability to modify device parameters.
Timeline
- CVE published
- CISA SSVC assessment
Exploitation: none · Automatable: no · Technical impact: total
- Record last updated
Affected versions
- Endress+Hauser Promag 10 with HART
fixed in 01.00.06
- Endress+Hauser Promag 10 with IO-Link
fixed in 01.00.02
- Endress+Hauser Promag 10 with Modbus
fixed in 01.00.06
- Endress+Hauser Promass 10 with HART
fixed in 01.00.06
- Endress+Hauser Promass 10 with IO-Link
fixed in 01.00.02
- Endress+Hauser Promass 10 with Modbus
fixed in 01.00.06