Skip to main content
← All vulnerabilities
CVE-2025-14097

Remote Code Execution Vulnerability in Radiometer Products

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub
  • Exploitation: none
  • Automatable: no
  • Technical impact: total

Summary

A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software.                                                                                                                                                                                                Other related CVE's are CVE-2025-14095 & CVE-2025-14096.                                                                                                      Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer.                                                                                                                                                                        Temporary work Around: If the network is not considered secure, please remove the analyzer from the network.                         Permanent solution: Customers should ensure the following: • The network is secure, and access follows best practices. Local Radiometer representatives will contact all affected customers to discuss a permanent solution.                                                      Exploit Status: Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.

Timeline

  1. CVE published
  2. CISA SSVC assessment

    Exploitation: none · Automatable: no · Technical impact: total

  3. Record last updated

Affected versions

  • Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers

    Version range not reported

  • Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX Analyzers

    Version range not reported

  • Radiometer Medical Aps AQT90 FLEX Analyzers

    Version range not reported

References 1

radiometer.com