Skip to main content
← All vulnerabilities
CVE-2025-14096

Credential Disclosure vulnerability in Radiometer Products

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub
  • Exploitation: none
  • Automatable: no
  • Technical impact: total

Summary

A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.

Timeline

  1. CVE published
  2. CISA SSVC assessment

    Exploitation: none · Automatable: no · Technical impact: total

  3. Record last updated

Affected versions

  • Radiometer Medical Aps ABL90 FLEX and ABL90 FLEX PLUS Analyzers

    Windows 7 Operating system

  • Radiometer Medical Aps AQT90 FLEX Analyzers

    Windows 7 Operating system

  • Radiometer Medical Aps ABL800 BASIC and ABL800 FLEX analyzers

    Windows 7 Operating system

References 1

radiometer.com