CVE-2025-0626
Hidden Functionality vulnerability in Contec Health CMS8000 Patient Monitor
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
CISA's decision points: whether it is exploited, whether attacks can be automated, and how much control an attacker gets. Assessed (SSVC 2.0.3).What do these mean? →
- Exploitation: none
- Automatable: no
- Technical impact: total
Summary
The "monitor" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device.
Timeline
- CVE published
- CISA SSVC assessment
Exploitation: none · Automatable: no · Technical impact: total
- Record last updated
Affected versions
- Contec Health CMS8000 Patient Monitor
All versions