Skip to main content
← All vulnerabilities
CVE-2024-48974

Life2000 Ventilator does not perform proper file integrity checks when adopting firmware updates

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub
  • Exploitation: none
  • Automatable: no
  • Technical impact: total

Summary

The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.

Timeline

  1. CVE published
  2. CISA SSVC assessment

    Exploitation: none · Automatable: no · Technical impact: total

  3. Record last updated

Affected versions

  • Baxter Life2000 Ventilation System

    Version range not reported

References 1

cisa.gov