CVE-2021-37166
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by rel
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
Summary
A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.
Timeline
- CVE published
- Record last updated
Affected versions
- n/a n/a
Version range not reported
References 4
swisslog-healthcare.com
- swisslog-healthcare.com
- en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20
- -/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37166-bulletin---gui-socket-denial-of-service.pdf?rev=05321b2af1064eb2a6d6e6bf77604c6b&hash=40A927FE1153AA980428C93B2EF7EB40