CVE-2021-37162
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software be
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
Summary
A buffer overflow issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. If an attacker sends a malformed UDP message, a buffer underflow occurs, leading to an out-of-bounds copy and possible remote code execution.
Timeline
- CVE published
- Record last updated
Affected versions
- n/a n/a
Version range not reported
References 4
swisslog-healthcare.com
- swisslog-healthcare.com
- en-us/customer-care/security-information/cve-disclosures#:~:text=CVE%20Disclosures%20%20%20%20Vulnerability%20Name%20%2C%20%20CVE-2021-37164%20%204%20more%20rows%20
- -/media/swisslog-healthcare/documents/customer-service/armis-documents/cve-2021-37162-bulletin---overflow-in-sccprocessmsg.pdf?rev=55a2a1d76199435688a8479970fc54bf&hash=4FDAB2F0EB319F0B773500669D67F3AD