Skip to main content
← All vulnerabilities
CVE-2021-37160

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of softwar

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • n/a n/a

    Version range not reported

References 4

swisslog-healthcare.com

armis.com