Skip to main content
← All vulnerabilities
CVE-2020-25183

Medtronic MyCareLink Smart Improper Authentication

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Medtronic MyCareLink Smart 25000 contains an authentication protocol vulnerability where the method used to authenticate between the MCL Smart Patient Reader and the Medtronic MyCareLink Smart mobile app is vulnerable to bypass. This vulnerability enables an attacker to use another mobile device or malicious application on the patient’s smartphone to authenticate to the patient’s Medtronic Smart Reader, fooling the device into believing it is communicating with the original Medtronic smart phone application when executed within range of Bluetooth communication.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Medtronic Smart Model 25000 Patient Reader

    All versions

References 2

global.medtronic.com

cisa.gov