CVE-2020-15486
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices.
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
Summary
An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.
Timeline
- CVE published
- Record last updated
Affected versions
- n/a n/a
Version range not reported