Skip to main content
← All vulnerabilities
CVE-2020-15486

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices.

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • n/a n/a

    Version range not reported

References 1

payatu.com