Skip to main content
← All vulnerabilities
CVE-2020-12048

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TL

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker with access to the network could observe sensitive treatment and prescription data sent between the Phoenix system and the Exalis tool.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • n/a Baxter Phoenix Hemodialysis Delivery System

    Version range not reported

References 1

us-cert.gov