Skip to main content
← All vulnerabilities
CVE-2020-10627

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Ins

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Insulet Omnipod Insulin Management System insulin pump product ID 19191 and 40160 is designed to communicate using a wireless RF with an Insulet manufactured Personal Diabetes Manager device. This wireless RF communication protocol does not properly implement authentication or authorization. An attacker with access to one of the affected insulin pump models may be able to modify and/or intercept data. This vulnerability could also allow attackers to change pump settings and control insulin delivery.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Insulet Omnipod Insulin Management System

    19191, 40160, ZXP425, ZXR425

References 2

us-cert.cisa.gov

myomnipod.com