Summary

In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism does not apply read protection, allowing for full read access of the RFID security mechanism data.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Medtronic Valleylab FT10 Energy Platform (VLFT10GEN)

    through 2.1.0; through 2.0.3

  • Medtronic Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)

    through 1.20.2