CVE-2019-13531
Medtronic Valleylab FT10 and LS10 Improper Authentication
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
Summary
In Medtronic Valleylab FT10 Energy Platform (VLFT10GEN) version 2.1.0 and lower and version 2.0.3 and lower, and Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States) version 1.20.2 and lower, the RFID security mechanism used for authentication between the FT10/LS10 Energy Platform and instruments can be bypassed, allowing for inauthentic instruments to connect to the generator.
Timeline
- CVE published
- Record last updated
Affected versions
- Medtronic Valleylab FT10 Energy Platform (VLFT10GEN)
through 2.1.0; through 2.0.3
- Medtronic Valleylab LS10 Energy Platform (VLLS10GEN—not available in the United States)
through 1.20.2