CVE-2018-8870

Medtronic MyCareLink Patient Monitor Use of Hard-coded Password

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains a hard-coded operating system password. An attacker with physical access can remove the case of the device, connect to the debug port, and use the password to gain privileged access to the operating system.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Medtronic 24950 MyCareLink Monitor

    All versions

  • Medtronic 24952 MyCareLink Monitor

    All versions

References 2

global.medtronic.com

ics-cert.us-cert.gov