Skip to main content
← All vulnerabilities
CVE-2018-19010

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions.

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Drager Infinity Delta, Infinity Delta, all versions, Delta XL, all versions, Kappa, all version, and Infinity Explorer C700, all versions. A malformed network packet may cause the monitor to reboot. By repeatedly sending the malformed network packet, an attacker may be able to disrupt patient monitoring by causing the monitor to repeatedly reboot until it falls back to default configuration and loses network connectivity.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • ICS-CERT Dräger Infinity Delta

    Version range not reported

References 2

ics-cert.us-cert.gov

securityfocus.com