In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser passwor

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser password and physical access can enter the superuser password that can be used to access and modify all settings on the device, as well as allow the user to reset existing passwords.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs

    All versions prior to May 2018.

References 3

ics-cert.us-cert.gov

usa.philips.com

securityfocus.com