Skip to main content
← All vulnerabilities
CVE-2018-14781

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Medtronic MMT- 508 - MiniMed pump

    All versions

  • Medtronic MMT – 511 pump Paradigm

    All versions

  • Medtronic MMT – 512 / MMT – 712 Paradigm x12

    All versions

  • Medtronic MMT – 515 / MMT – 715 Paradigm x15

    All versions

  • Medtronic MMT – 522 / MMT – 722 Paradigm REAL-TIME

    All versions

  • Medtronic MMT – 522(K) / MMT – 722(K) Paradigm REAL-TIME

    All versions

  • Medtronic MMT – 523 / MMT – 723 Paradigm Revel

    All versions

  • Medtronic MMT – 523(K) / MMT – 723(K) Paradigm

    All versions

  • Medtronic MMT – 554 / MMT – 754 MiniMed Veo

    All versions

  • Medtronic MMT – 551 / MMT – 751 MiniMed 530G

    All versions

References 3

global.medtronic.com

ics-cert.us-cert.gov

securityfocus.com