CVE-2017-9656
The backend database of the Philips DoseWise Portal application versions 1.1.7.333 and 2.1.1.3069 uses hard-coded credentials for a database
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
Summary
The backend database of the Philips DoseWise Portal application versions 1.1.7.333 and 2.1.1.3069 uses hard-coded credentials for a database account with privileges that can affect confidentiality, integrity, and availability of the database. For an attacker to exploit this vulnerability, elevated privileges are first required for an attacker to access the web application backend system files that contain the hard-coded credentials. Successful exploitation may allow a remote attacker to gain access to the database of the DWP application, which contains PHI. CVSS v3 base score: 9.1, CVSS vector string: AV:N/AC:L/PR:H/UI:N/S:C/C:H/I:H/A:H.
Timeline
- CVE published
- Record last updated
Affected versions
- Philips DoseWise Portal
1.1.7.333, 2.1.1.3069