Skip to main content
← All vulnerabilities
CVE-2017-5149

An issue was discovered in St.

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • n/a St. Jude Merlin@home Transmitter before 8.2.2

    St. Jude Merlin@home Transmitter before 8.2.2

References 2

securityfocus.com