CVE-2017-12726

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump