CVE-2017-12722

An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • n/a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump

    Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump