Skip to main content
← All vulnerabilities
CVE-2015-1012

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System.

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Hospira LifeCare PCA Infusion System

    through 5.0

References 1

ics-cert.us-cert.gov