Skip to main content
← All vulnerabilities
CVE-2014-5406

Hospira LifeCare PCA Infusion System

Published
Last modified
published
Not listed in CISA KEV
None on Exploit-DB, Metasploit or GitHub

Summary

The Hospira LifeCare PCA Infusion System before 7.0 does not validate network traffic associated with sending a (1) drug library, (2) software update, or (3) configuration change, which allows remote attackers to modify settings or medication data via packets on the (a) TELNET, (b) HTTP, (c) HTTPS, or (d) UPNP port. NOTE: this issue might overlap CVE-2015-3459.

Timeline

  1. CVE published
  2. Record last updated

Affected versions

  • Hospira LifeCare PCA Infusion System

    through 5.0

References 4

cisa.gov

fda.gov

xs-sniper.com

GitHub · cisagov/CSAF