Unspecified vulnerability in Medtronic Paradigm wireless insulin pump 512, 522, 712, and 722 allows remote attackers to modify the delivery
- Published
- Last modified
- Where the record stands: published, rejected by its publisher, or superseded by another id.What do these mean? →
- published
- Whether CISA lists it in its Known Exploited Vulnerabilities catalog: proof that attackers are using it.What do these mean? →
- Not listed in CISA KEV
- Whether working attack code is public (Exploit-DB, Metasploit, Nuclei or GitHub). Public code makes attacks easier.What do these mean? →
- None on Exploit-DB, Metasploit or GitHub
Summary
Unspecified vulnerability in Medtronic Paradigm wireless insulin pump 512, 522, 712, and 722 allows remote attackers to modify the delivery of an insulin bolus dose and cause a denial of service (adverse human health effects) via unspecified vectors involving wireless communications and knowledge of the device's serial number, as demonstrated by Jerome Radcliffe at the Black Hat USA conference in August 2011. NOTE: the vendor has disputed the severity of this issue, saying "we believe the risk of deliberate, malicious, or unauthorized manipulation of medical devices is extremely low... we strongly believe it would be extremely difficult for a third-party to wirelessly tamper with your insulin pump... you would be able to detect tones on the insulin pump that weren't intentionally programmed and could intervene accordingly."
Timeline
- CVE published
- Record last updated
Affected versions
- n/a n/a
Version range not reported